A-A-51266
The package shall consist of forty-eight (48) filled cartridges: twenty-four (24) 19-gauge disposable dispensing needles, twenty-four (24) 21-gauge disposable dispensing needles, and a set of adequate and accurate instructions for use. Specific instructions shall be furnished for enlarging the hole in the nose of the syringe in case the cannula of the 19-gauge needle does not easily pass through the hole.
Workmanship. The impression material shall be clean and free from defects which may affect proper functioning or serviceability.
Unit. Package (PG). One package containing forty-eight cartridges, as specified, constitutes one unit.
Quality Assurance Provisions.
Responsibility for inspection. Unless otherwise specified in the contract or purchase order, the contractor is responsible for the performance of all inspection requirements as specified herein. Except as otherwise specified in the contact or purchase order, the contractor may use his own or any facilities suitable for the performance of the inspection requirements specified herein, unless disapproved by the Government. The Government reserves the right to perform any of the inspections set forth in the specification where such inspections are deemed necessary to assure supplies and services conform to prescribed requirements.
Records. Records of examinations and tests performed by or for the contractor shall be maintained by the contractor and made available to the Government upon the Government’s request, at any time, or from time to time, during the performance of the contract and for a period of three years after delivery of the supplies to which such records relate.
Inspection. Inspection, as used herein, is defined as both examination (such as visual or auditory investigation without the use of special laboratory appliances or procedures) and testing (determination by technical means of physical and chemical properties) of the item.
Sampling for inspection. Sampling for inspection shall be conducted in accordance with MIL-STD-105. Where feasible, the same sample shall be used for the determination of two or more inspection characteristics. The unit of product for inspection purposes shall be one unit package. Unless otherwise specified, sampling for examination shall be conducted at inspection level II, AQL 2.5, and sampling for tests (when applicable) shall be conducted on one unit package from each lot or batch number. Failure of any test shall cause rejection of the lot or batch number.