A-A-54529
The system shall comply with the Code of Federal Regulations (CFR) 47, Section 18, and applicable UL Standards, which shall be evidenced by UL listing, CSA approval, City of Loa Angeles approval, or testing by an independent laboratory.
The ultrasonic cleaning system shall be capable of operating over the full tolerance range of one or more of the following line voltages and frequencies as specified by the customer requisition:
* 115 VAC plus or minus 10 percent, 90 Hz plus or minus 2 Hz
* 115 VAC plus or minus 10 percent, 60 Hz plus or minus 2 Hz
* 230 VAC plus or minus 10 percent, 50 Hz plus or minus 2 Hz
* 230 VAC plus or minus 10 percent, 60 Hz plus or minus 2 Hz
Dual voltage and hertz on all units are desirable but not mandatory.
A three-wire power cord and plug, which has been approved by a nationally recognized testing laboratory as suitable for hospital environments, shall be supplied.
Operational and Service Data Manuals shall be furnished as follows:
a. Information Content:
The contractor shall furnish manuals, handbooks, and/or brochures containing complete operation, installation, and service/maintenance instructions (including pictures or illustrations, as necessary) with complete schematics and wiring diagrams. All manuals, handbooks, and/or brochures shall be written in the English language, and all schematics and wiring diagrams shall use American electrical and electronics symbols. The manuals shall include electrical data and connection diagrams for all applicable utilities. The instructions shall also contain a complete list of all replaceable parts showing part number, name, and quantity required. All service notes, service memos, etc. pertaining to the equipment and issued subsequent to the printing of the original instructions shall also be included. Three instructions shall be, as a minimum, those furnished to service engineers (or servicemen) who normally install and service the equipment for the company or distributor. When the system being procured includes, or will operate