A-A-53939
Sampling for examination. Sampling for examination shall be conducted in accordance with MIL-STD-105, at an inspection level of S-4, with an AQL of 2.5 (percent defective). The unit of product for examination shall be one point for each package sampled.
Examination. The points shall be examined to determine compliance with all requirements specified herein. Points shall be examined for defects, including but not limited to, those listed below.
Defects
Points are not clean.
Style and design not as specified.
Color of points not as specified.
Tapered length surface not smooth rolled finish.
Laminations cracks or voids present.
Twisted or curled sections present.
Size not as specified.
Sampling for tests. Sampling for tests shall be conducted in accordance with MIL-STD-105, at an inspection level of S-1, with an AQL of 1.0 (percent defective). The unit of product for tests purposes shall be one point from each package sampled.
Tests. Tests shall be conducted to determine compliance with specification requirements. Where feasible, the same sample shall be used for the determination of two or more test characteristics. There shall be two determinations per lot or batch number. Each determination shall be the average of specimens when averaging is specified in testing procedure.
Dimensions. The points shall be tested for dimensions in accordance with ANSI/ADA No. 57.